Hybono HY-CS1753D is a general-purpose digital pH sensor. The industrial electrode adopts the world's most advanced solid dielectric and large area PTFE liquid junction. Not easy to block, easy to maintain, built-in temperature sensor for automatic temperature compensation, support RS485 Modbus digital output. It is ideal for long-term online monitoring.
Overview
Designed for strong acid, strong base, wastewater, and chemical process.
Easy to connect to PLC, DCS, industrial control computers, general-purpose controllers, paperless recording instruments or touch screens, and other third-party devices.
HY-CS1753D pH electrode adopts the most advanced solid dielectric in the world and large-area PTFE liquid junction. Not easy to block, easy to maintain.
The long-distance reference diffusion path greatly prolongs the service life of the electrode in harsh environments. With a built-in temperature sensor (NTC10K, Pt100, Pt1000, etc. can be selected according to user requirements) and a wide temperature range, it can be used in explosion-proof areas.
The newly designed glass bulb increases the bulb area, prevents the generation of interfering bubbles in the internal buffer, and makes the measurement more reliable. Adopt PPS/PC shell, upper and lower 3/4NPT pipe thread, easy to install, no need for sheath, and low installation cost. The electrode is integrated with pH, reference, solution grounding, and temperature compensation.
The electrode adopts a high-quality low-noise cable, which can make the signal output longer than 20 meters without interference.
The electrode is made of ultra-bottom impedance-sensitive glass film, and it also has the characteristics of fast response, accurate measurement, good stability, and not easy to hydrolyze in the case of low conductivity and high purity water.
Technical Parameters | |
Model No. | HY-CS1753D |
Power/Outlet | 9~36VDC/RS485 MODBUS RTU or 4~20mA |
Measure material | Glass/silver+ silver chloride |
Housing material | PP |
Waterproof grade | IP68 |
Measurement range | 0-14pH |
Accuracy | ±0.05pH |
Pressure resistance | ≤0.6Mpa |
Temperature compensation | NTC10K |
Temperature range | 0-80ºC |
Calibration | Sample calibration, standard liquid calibration |
Connection methods | 4 core cable |
Cable length | Standard 10m cable, can be extended to 100m |
Installation thread | NPT3/4'' |
Application | General application, industrial water, sewage, river, lake,etc |

-